1-U2B/5KN 可靠的拉伸力和压缩力传感器

¥10,998.00

侧向力补偿
集成侧向力补偿
不锈钢,防护等级 IP67
额定力 500 N… 200 kN
可根据要求配置不同的电缆长度、连接器附件、集成放大器(0 … 10 V,4 … 20 mA 或 IO-Link)以及 TEDS

描述

Specifications without integrated amplifier
Type U2B without integrated amplifier
Nominal (rated) force F nom N 500
kN 1 2 5 10 20 50 100 200
Accuracy
HBK accuracy class 0.2 0.1
Relative reproducibility and
repeatability errors in unchanged
mounting position
brg % 0.1
Relative reversibility error
(hysteresis) at 0.5 * F nom
V0.5 % 0.2 0.15
Non-linearity dlin % 0.2 0.1
Relative zero point return vw0 % 0.05
Relative creep over 30 min dcr, F+E % 0.06
Effect of the bending moment
at 10% F nom * 10mm
dMb % 0.05
Effect of lateral forces at 10% of
F nom
d Q % 0.1
Temperature coefficient of
sensitivity
TC S %/10K 0.1
Temperature coefficient of zero
signal
TC0 %/10K 0.1 0.05
Rated electrical output
Rated output (nominal) C nom mV/V 2
Relative zero signal error ds,0 % 1
Characteristic curve deviation,
tension
dc % 0.2
Characteristic curve deviation,
tension/compression
dzd % 1.5 0.5
Input resistance Re Ω >345
Output resistance Ra Ω 300 … 400
Insulation resistance Riso > 2
Operating range of the excitation
voltage
BU,G V 0.5 … 12
Reference excitation voltage U ref V 5
Connection 6-wire circuit
Temperature
Reference temperature T ref °C [°F] +23 [73.4]
Nominal (rated) temperature range BT,nom °C [°F] ‐10 … +70 [14 … 158]
Operating temperature range BT,G °C [°F] ‐30 … +85 [-22 … +185]
Storage temperature range BT,S °C [°F] ‐50 … +85 [-58 … +185]
Characteristic mechanical quantities
Maximum operating force FG % of
F nom
130 150
Force limit FL 130 150
Breaking force FB > 300
Torque limit MG
max
Nm 46.5 63 63 60 108 340 620 2430 5125
Bending moment limit Mb
max
2.9 12.8 19 24 49 223 380 1463 2880
Static lateral force limit F Q    % of
F nom
25 52 36 18 25 35 19 25 19
Nominal (rated) displacement s nom mm 0.058 0.056 0.048 0.047 0.047 0.065 0.082 0.09 0.12
Natural frequency fG kHz 4 6 8.7 14 17.5 8 8.5 6 5.6
Permissible oscillation stress frb    % of
F nom
100 160
Stiffness cax  105 N/
mm
   0.08
6
0.18 0.42 1.06 2.13 3.08 6.1 11.1 16.67
General information
Degree of protection in accordance with EN 60529 IP67
Spring element material Stainless steel
Measuring point protection Hermetically-welded measuring body
Cable (only with “fixed cable” option) 6‐wire, polyethylene insulated
Cable length (standard version) m 3 6 12
Cable length (to customer requirements) See “U2B versions and ordering numbers”
Weight m kg 0.8 2.9 4.3 10.7 15.9
m lbs 1.76 6.4 9.48 23.6 35.05
Mechanical shock resistance as per IEC 60068-2-6
Number n 1000
Duration ms 3
Acceleration m/s2 637
Vibrational stress as per IEC 60068-2-27
Frequency range Hz 5 … 65
Duration min 30
Acceleration m/s2 150
1)  Test condition: 1 m water column, 0.5 hours; connected with cable if version with M12 male connectorselected

 

 

Specifications with integrated amplifier VA1 (0…10 V) and VA2 (4…20 mA)
Type U2B with integrated VA1 and VA2 amplifier
Nominal (rated) force F nom N 500
kN 1 2 5 10 20 50 100 200
Accuracy
HBK accuracy class 0.2 0.1
Relative reproducibility and
repeatability errors in unchanged
mounting position
brg % 0.1
Relative reversibility error
(hysteresis) at 0.5 * F nom
V0.5 % 0.2 0.15
Non-linearity dlin % 0.2 0.1
Relative zero point return vw0 % 0.05
Relative creep over 30 min dcr, F+E % 0.06
Effect of the bending moment
at 10% F nom * 10 mm
dMb % 0.05
Effect of lateral forces at 10% of
F nom
d Q % 0.1
Temperature coefficient of
sensitivity
TC S %/10K 0.1
Temperature coefficient of zero
signal
TC0 %/10K 0.1 0.05
Rated electrical output VA1 (voltage output)
Output signal V 0 … 10
Rated output (nominal) V 5
Sensitivity tolerance V ±0.1
Zero signal V 5
Rated output variation for tension/
pressure
dzd % 1.5 0.05
Output signal range V -0.3 … 11
Cut-off frequency (-3 db) fG kHz 2
Nominal (rated) supply voltage U ref V 24
Operating range of the supply
voltage
Bu,gt V 19 … 30
Max. current consumption mA 15
Electrical connection M12 connector, 8-pin, A-coded
Rated electrical output VA2 (voltage output)
Output signal mA 4 … 20
Rated output (nominal) mA 8
Sensitivity tolerance mA ±0.16
Zero signal mA 12
Rated output variation for tension/
pressure
dzd % 1.5 0.05
Output signal range mA 3 … 21
Cut-off frequency (-3 db) fG kHz 2
Nominal (rated) supply voltage U ref V 24
Operating range of the supply
voltage
Bu,gt V 19 … 30
Max. current consumption mA 30
Electrical connection M12 connector, 8-pin, A-coded
B00482 05 E00 04   08.01.2026                                                                                                                                                                               9
Type U2B with integrated VA1 and VA2 amplifier
Nominal (rated) force N 500
kN 1 2 5 10 20 50 100 200
Temperature
Reference temperature T ref °C [°F] +23 [73.4]
Nominal (rated) temperature range BT,nom °C [°F] ‐10 … +50 [14 … 122]
Operating temperature range BT,G °C [°F] ‐20 … +60 [-4 … +140]
Storage temperature range BT,S °C [°F] ‐25 … +85 [-77 … +185]
Characteristic mechanical quantities
Maximum operating force FG     % of
F nom
130 150
Force limit FL 130 150
Breaking force FB > 300
Torque limit MG max Nm 46.5 63 63 60 108 340 620 2430 5125
Bending moment limit Mb max 2.9 12.8 19 24 49 223 380 1463 2880
Static lateral force limit Fq    % of
F nom
25 52 36 18 25 35 19 25 19
Nominal (rated) displacement s nom mm 0.058 0.056 0.048 0.047 0.047 0.065 0.082 0.09 0.12
Natural frequency fG kHz 4 6 8.7 14 17.5 8 8.5 6 5.6
Permissible oscillation stress frb    % of
F nom
100 160
Stiffness cax  105 N/
mm
0.086 0.18 0.42 1.06 2.13 3.08 6.1 11.1 16.67
General information
Degree of protection in accordance with EN 60529 IP67
Spring element material Stainless steel
Material of permanently installed amplifier housing Stainless steel
Measuring point protection Hermetically-welded measuring body
Weight m kg 0.8 2.9 4.3 10.7 15.9
m lbs 1.76 6.4 9.48 23.6 35.05
Mechanical shock resistance as per IEC 60068-2-6
Number n 1000
Duration ms 3
Acceleration m/s2 637
Vibrational stress as per IEC 60068-2-27
Frequency range Hz 5 … 65
Duration min 30
Acceleration m/s2 150
2)  Test condition: 1 m water column, 0.5 hours; connected with cable if version with M12 male connectorselected

 

 

Specifications with integrated VAIO amplifier
Type U2B with integrated VAIO amplifier
Nominal (rated) force F nom N 500
kN 1 2 5 10 20 50 100 200
Accuracy
HBK accuracy class 0.2 0.1
Relative reproducibility and
repeatability errors in unchanged
mounting position
brg % 0.1
Relative reversibility error
(hysteresis) at 0.5 * F nom
V0.5 % 0.2 0.15
Non-linearity dlin % 0.03
Relative zero point return vw0 % 0.05
Relative creep over 30 min dcr, F+E % 0.06
Effect of the bending moment
at 10% F nom * 10 mm
dMb % 0.05
Effect of lateral forces at 10% of
F nom
d Q % 0.1
Temperature coefficient of
sensitivity
TC S %/10K 0.1
Temperature coefficient of zero
signal
TC0 %/10K 0.03
VAIO electrical characteristics
Output signal, interface COM3, to IO-Link standard, class A
Min. cycle time ms 0.9
Sample rate (internal) S/s 40000
Cut-off frequency (-3 db) fG kHz 4
Nominal (rated) supply voltage U ref V 24
Operating range of the supply
voltage
Bu,gt V 19 … 30
Maximum power consumption mW 3200
Noise ppm of
nomi-
nal
force
With Bessel filter 1Hz: 25
With Bessel filter 10 Hz: 63
With Bessel filter 100 Hz: 195
With Bessel filter 200 Hz: 275
Without filter: 3020
Low-pass filter Freely adjustable cut-off frequency, Bessel or Butterworth
characteristic, 6th order
Variation in relative characteristic
value for tension/pressure
dzd % 0.03
Device functions
Limit value switches 2 limit value switches. Invertible, freely adjustable hysteresis.
Output via process data or digital output
Digital I/O According to IO-Link Smart Sensor Profile, 1 permanently available
digital output; 1 output can be set to data output. Measurement is
then not possible
Slave pointer function Yes
Peak value memory Yes
Peak-to-peak memory Yes
Warning functions Warning on exceeding nominal (rated) force/maximum operating
force; Nominal (rated) temperature/maximum operating force
Temperature
Reference temperature T ref °C [°F] +23 [73.4]
B00482 05 E00 04   08.01.2026                                                                                                                                                                             11
Type U2B with integrated VAIO amplifier
Nominal (rated) force N 500
kN 1 2 5 10 20 50 100 200
Nominal (rated) temperature range BT,nom °C [°F] ‐10 … +50 [14 … 122]
Operating temperature range BT,G °C [°F] ‐10 … +60 [14 … +140]
Storage temperature range BT,S °C [°F] ‐25 … +85 [-77 … +185]
Characteristic mechanical quantities
Maximum operating force FG     % of
F nom
130 150
Force limit FL 130 150
Breaking force FB > 300
Torque limit MG max Nm 46.5 63 63 60 108 340 620 2430 5125
Bending moment limit Mb max 2.9 12.8 19 24 49 223 380 1463 2880
Static lateral force limit Fq    % of
F nom
25 52 36 18 25 35 19 25 19
Nominal (rated) displacement s nom mm 0.058 0.056 0.048 0.047 0.047 0.065 0.082 0.09 0.12
Natural frequency fG kHz 4 6 8.7 14 17.5 8 8.5 6 5.6
Permissible oscillation stress frb    % of
F nom
100 160
Stiffness cax  105 N/
mm
0.086 0.18 0.42 1.06 2.13 3.08 6.1 11.1 16.67
General information
Degree of protection in accordance with EN 60529 IP67
Spring element material Stainless steel
Material of permanently installed amplifier housing Stainless steel
Measuring point protection Hermetically-welded measuring body
Weight m kg 0.8 2.9 4.3 10.7 15.9
m lbs 1.76 6.4 9.48 23.6 35.05
Mechanical shock resistance as per IEC 60068-2-6
Number n 1000
Duration ms 3
Acceleration m/s2 637
Vibrational stress as per IEC 60068-2-27
Frequency range Hz 5 … 65
Duration min 30
Acceleration m/s2 150
3)  Test condition: 1 m water column, 0.5 hours; connected with cable if version with M12 male connectorselected